ISO-80 |
 |
 |
| |
The
improved ISO-80 provides low noise AC coupled
amplification and offers excellent recording performance
for monitoring extracellular nerve action potentials in
vitro and in living animals |
|
|

|
 |
 |
 |
 |
The
ISO-80 is provided with a remote headstage (1 m cable)
which incorporates an electrode impedance test function
and a constant current stimulator. The constant current
stimulator can be used for cell marking, stimulation or
electrode cleaning. Typical applications include measuring
EMG, EEG, extracellular and action potentials in vitro or
in vivo. The ISO-80 system is DC isolated from the subject
ground and employs state of the art electro-magnetic
shielding for improved noise rejection. The amplifier
employs both high pass and low pass filtering with gain
from 100 to 10,000. The lowest low-pass setting is 5Hz and
the upper pass band is 10Khz.
The ISO-80 Start-up Kit now contains all the below listed
accessories needed for basic metal electrode
electrophysiology research:
CBL102 Cable, BNC-to-3.5mm plug, 6 ft (2m) (two)
5469 Adapter, mini-banana to 0.031 skt. (two)
13388 Adapter, mini-banana to 2mm skt. (two)
3294 Cable, ground clip to wire, 3 ft
2033 Mini-banana plug, black
2034 Mini-banana plug, red
2035 Mini-banana plug solderable turrent (two)
EP1 Ag/AgCl pellet (70 mm wire) 1mm diam x 2.5 mm long
M3301EH Electrode Holder, 14cm (two)
5470 0.031-inch jack on 12-inch wire (package of 4)
APPLICATIONS:


|
 |
| |
|
|
| |
| SPECIFICATIONS |
| Input
Resistance |
>1011
Ohms, Common Mode and differential |
| Input
Leakage Current |
50
picoamperes, max. |
| Amplification |
x102,
x103, x104< |
| Common
Mode Rejection Ratio |
100
dB typ. @ 50/60 Hz |
| Equivalent
Noise Signal Input |
0.4
microvolts rms (0.1-100 Hz)
2.0 microvolts rms (1 Hz - 10 kHz) |
| Filter
Settings |
|
Low frequency
|
5,
10, 300 Hz |
|
High frequency
|
100
Hz, 1, 3, 10 kHz |
| Max.
Output Voltage Swing |
+8
volts |
| Electrode
Impedance Range |
100
kOhm - 10 MOhm @ 300 Hz |
| Stimulation
Current |
0
to ±20 micro amperes (constant current) |
| Maximum
Electrode Voltage |
±40
volts |
| Display |
3
_ LCD |
| Battery
Test |
1
second audible tone |
| Power |
2
nine-volt alkaline batteries, supplied |
|
MODELS |
| ISO 80 |
Isolated Bio-amplifier w/
active probe (ISO-80P) |
 |
| OPTIONAL
ACCESSORIES |
| ISO 80P |
Replacement ISO-80 Probe |
 |
|
|
 |
| Cables and Connectors
| Manual  |
 |
 |
 |
 |
 |
 |
 |
|
 |