The IITC plantar and tail flick combination system is
the only commercially available system of its kind. No
need to purchase two separate systems, our combinations
system enables researchers to perform both tests with
this unique design. Testing properties of narcotic and
stronger non-narcotic drugs on unrestrained mice and
rats in plantar mode.
The IITC Plantar Analgesia Meter for paw stimulation
can be used on 12 mice, 6 rats and other animals (cats,
rabbits) unrestrained when testing for narcotic drugs.
Experiments are easy to perform, simply slide the test
head under test subject, align the heat source via our
exclusive guide light (idle state) by the attached,
adjustable, angled mirror on test head to test subject
and perform tests.
Start, stop and reset of tests is done by either the
push button on test head, front panel controls or
supplied footswitch.
With our exclusive “idle state” of the heat source
one is assured of exact placement of the heat source.
Tests are performed on the plantar surface by a focused,
radiant heat light source. The light beam is focused to
the top of the glass and creates a 4X6mm intense spot on
the paw.
Our method of visible light is superior to the
infrared light method, since a user will not always know
exactly where/when the test will start, causing false
starts and conditioning of animals. A built-in timer
displays reaction time in .01 second increments.
The height of the glass is adjustable allowing the
focal point of the light to be adjusted to stimulate
other body parts if needed. A Humane cut-off timer is
standard. The unit can be preset for a cut-off period,
at the end of which the heat source shuts off
automatically if the animal has not responded, avoiding
tissue damage. Our test head does not emit , clicking or
whining noises that cause preconditioning of animals
prior to test as with other systems on the market.
Our exclusive heated glass option eliminates the
glass acting as a heat sink. This feature will eliminate
the reaction time of the animals to be altered due to
the fact that they are placed on cold glass allowing the
possibility of delayed response time to the applied heat
source. Built-in heating elements in the glass which are
digitally controlled for consistent temperatures.
Temperature readout and adjustment from 30-40 degrees c.
for the heated base is controlled on the heated base.
 |
 |
| Tail Temperature
Option |
Part# 433 Animal
enclosure |
Three individual, acrylic animal enclosures are
supplied. Each enclosure allows for 2 rats/4 mice
allowing testing of up to 6 rats/12mice total.
Enclosures have strong magnets to prevent animals from
possible escape., individual lids, black dividers in
between test subjects.
Tail Flick mode - the test head is attached to the
electronic gear and the heat source now sits over a tail
groove that has a sensor that detects the when the
animal flicks it tail. System records and displays time
and temperature of test. A humane cut-off timer that can
be set by user, preventing tail tissue damage if test
subject has not reacted within preset time.
TAIL TEMPERATURE OPTION (T)
Measurement of tail temperature at all times.
Trigger temperature is automatic at the start of
every test.
This optional feature allows automatic preheating of
the tail. To ensure experiment protocols to be precisely
controlled and repeated. This feature allows the user to
SET the tail temperature at start of each test is to be
started at up to 75 degrees Celsius. Once that preset
temperature is reached the test and the timer starts
automatically and after the tail flicks the timer and
light stop and the result is displayed. Readout will now
show in repeatable sequence the start and end
temperatures as well as test time. We recommend animals
be restrained with one of the IITC restrainers.
This unit solves the problem associated with “tail
temperature prior to and at the end of testing.” See
article: KJELL HOLE AND ARNE TJOLSEN: The Tail-Flick
Formalin Test in Rodents: Changes in Skin Temperature as
a Confounding Factor. Pain 53 (1993) pages. 247-254
Elsevier.
Two year warranty
120/240V
CE
-
Tail Temperature optional (T)
-
Precise Programmable digital
control
-
Humane cut-off feature, set by
user
-
Adjustable beam intensity in 1%
increments
-
Reaction is detected
automatically (Tail Flick set-up)
-
Manual override of all timer
functions
-
All Functions and parameters
entered via key-pad
-
Heated glass option only from
IITC
-
Footswitch - RS232 - Printer
output (standard)
-
Software - USB - Printer
(optional)
|