Tips
consist of 2 mm diameter plastic tubes sealed at one end
with an ion-sensitive membrane. After filling with
electrolyte solution, the user inserts the tube into the
holder and connects it to a pH meter. Tips and holders are
interchangeable, so one tip may be replaced with another
sensitive to a different ion. Replacing a tip takes less
than a minute. Electrode tips normally last several
months, when stored properly in saline solution. When
replacement is necessary, only the tip need be replaced.
Kwik-Tip
electrodes are available separately and as kits -- Each
"KWIK" Electrode Holder kit includes a reusable
holder and three removable tips. In addition to a 4-foot
BNC cable and an electrolyte filling syringe;
"TIP" Electrode Kits contain three electrode
tips for a specific ion. A separate reference electrode,
such as WPI's Dri-Ref, is also required.
If
your pH meter requires a US Standard connector, also order
Part No. 3508 (BNC-to-US pH Standard adapter).
|
TIP
ELECTRODE SPECIFICATIONS |
|
Part No.
|
Electrode
|
Color Code
|
Recommended Filling
Solution
|
Min.Slope / Decade
|
Concentration Range
|
Selectivity
Coefficients (-log)
|
|
TIPCA
|
Calcium
|
Green
|
0.1 M CaCl2
|
28 mV
|
0.1 M - 10-6.75
M
|
Na+ 5.5, K+ 5.4, Mg++ 4.9
|
|
TIPH
|
Hydrogen
|
Orange
|
1 M Citric Acid,0.01 M
NaCl, pH 5.6
|
54 mV
|
pH 5.0 - 12
|
Na+ 10.4, K+ 9.8, Ca++
11.1
|
|
TIPK
|
Potassium
|
Yellow
|
0.1 M KCl
|
54 mV
|
0.1 M - 10-4.5
M
|
Na+ 4.0, Ca++ 3.9, Mg++
3.0
|
| TIPTPP |
TPP+ |
Purple |
10mM TPP+ |
54 mV |
0.001 M - 10-4
M |
K+ 6.0 |
|