 Nanoliter Injector

Email to Colleague  

  WPI’s microprocessor-controlled Nanoliter 2000 uses direct piston displacement.   

By either pushing the injection button on the control box or pressing on the optional footswitch, a calibrated volume will be smoothly injected. The process is quiet and vibration free. Capillary filling and injection speeds are 23 nL/sec and 46 nL/sec (emptying speeds are 92 nL/sec and 230 nL/sec). Maximum fluid ejection is 5 µL. Each unit comes with sufficient glass to pull at least 300 tips. Glass is 1.14 mm O.D. (nominal) and 0.5 mm I.D.
By setting the DIP switch, the injection volume can be changed from 2.3 to 69.0 nL in 16 steps. Up to 100 injections may be triggered per filling. Since the volume of a normal Xenopus oocyte is about 500 nL, the instrument has the capability to inject from less than 1% to over 10% of the total volume of the oocyte in one preset step increment.
Included: 1 vial 3.5-in. capillaries (100); 1 vial 7-in. capillaries (100); replacement “O” rings; Allen wrench; MicroFil™ MF34G backfilling needle; and two sample µTip™ pre-pulled micropipettes.

WPI recommends the following equipment for a complete system,: PZMIV stereo microscope MMJ joystick micromanipulator M10 magnetic basePUL-100 micropipette puller or µTip™ micropipettes.
 

  Smart Controller

Micro4, an optional microprocessor-based controller, can provide an “intelligent” and easy-to-use interface to up to four Nanoliter Injectors. Operating parameters are set with the membrane keypad and LCD display. From the keypad the user can set pump to infusion or withdrawal mode, enter the volume to be infused or withdrawn, and rate of delivery, as well as synchronize the starting and stopping of a combination of Injectors. User parameters can be stored in the device’s “non-volatile” memory for instant recall when the unit is powered on. An optional footswitch can be plugged into a connector on the rear of the controller for “hands free” start / stop operation. An RS-232 port on the rear of the controller can be used to connect it to a computer for use with computer control programs.
  

 
 

Citations

1. Kay, B., and Peng, H.. "Xenopus laevis: Practical Uses in Cell and Molecular Biology". Methods in Cell Biology. Vol. 36. Academic Press
 
 

 

 
 

 

SPECIFICATIONS

INJECTION VOLUME Variable
REMOTE CONTROL Yes
GLASS OD 1.19 mm
GLASS ID 0.5 mm
INJECTION SPEED 23 nl/sec
FILL SPEED 23 nl/sec
EMPTY SPEED 46 nl/sec
VARIABLE VOLUME RANGE 2.3 - 73.6 nl
SMALLEST VOLUME 4.6 nl
TO CHANGE VOLUME Set switch
INJECTIONS PER FILLING, MAX. 100 injections
SHIPPING WEIGHT 3 lb. (1.1 kg)

MODELS

UMP2-1

UltraMicroPump II (one) and Micro4 Controller Add to Quote Basket
B203XVY Nanoliter 2000 (120 V, U.S. plug) Add to Quote Basket
B203XVZ Nanoliter 2000 (240 V, Continental plug) Add to Quote Basket
B203XVB Nanoliter 2000 (240 V, British plug) Add to Quote Basket
B203MC4 Nanoliter 2000 & Micro4 Controller Add to Quote Basket

OPTIONAL ACCESSORIES
 
13550 Volume Reduction Chip Add to Quote Basket
13142 Nanoliter Injector Footswitch Add to Quote Basket
4878 Replacement 3.5-in. glass capillaries (300) Add to Quote Basket
4879 Replacement 7-in. glass capillaries (300) Add to Quote Basket
TIP10XV119 Micropipettes for Nanoliter Injector (10) Add to Quote Basket
MICRO4 Micro4 Controller, Four-Channel Add to Quote Basket
40400 Adapter for Micro1 Add to Quote Basket
5340 Spare Parts Kit ( includes MicroFil MF34G, displacement piston, five O-ring sets ) Add to Quote Basket
500778 Nanoliter Ijector Universal Adapter Add to Quote Basket
 
| Manual 