Timing
Pulse
interval and width are set with single-turn continuously
variable controls from 5 ms to 5.5 s in three ranges.
Pulse width is continuously variable from 50 µs to 550 ms
in four ranges.
Modes
of operation
In
continuous mode, Isostim generates continuous square
waves. In gated or external sync modes, externally applied
pulses can generate trains or single events. Single pulses
of finite duration can be produced using a push-button on
the instrument's front panel. External/DC mode converts
Isostim to a passive stimulus isolator.
Dual
tone audible alarm
A
tone sounds when an open circuit is detected or when
system compliance is reached. A second tone, which sounds
when a signal is applied to the input, can only be heard
if the batteries have sufficient charge to operate the
isolator. A violation light advises when pulse width
exceeds the interval.
Current
delivery
Stimulus
currents up to 10 mA can be set on the front panel
with a control knob and a two-position range switch.
Output current is load-independent.
Power
Isostim
model A320D is powered by readily obtainable 9-volt
alkaline batteries (included). Under average use these
will last several months before replacement is required.
The rechargeable A320R is supplied with a nickel cadmium
battery stack which provides 10-12 hours of operation
before recharge is required. The A362 Battery Charger must
be used with the A320R.
|