Model
A395generates
an output current of arbitrary wave shape; DC, AC, pulse,
and combinations thereof. Battery operated, and
photoelectrically-isolated from the input voltage drive,
the instrument regenerates as output currents the original
waveforms provided by your D/A converter or signal
generator.
Current
Delivery -
Ten volts
input produces the maximum output current for the current
range selected, i.e., 100 µA, 1 mA, or 10 mA.
Front panel controls allow DC current to be generated.
Externally applied signals can be superimposed
simultaneously. Warning lamps indicate open circuit or
excessive current conditions.
Digital
Meter - Measures
DC or average output current.
Overload
Lamps - Indicate
when output voltage has reached positive or negative
compliance voltage limit.
|
 |
| |
|
|
| |
| A395
SPECIFICATIONS |
| OUTPUT
CURRENT, Imax |
3
ranges: 100µA, 1 mA, and 10 mA |
| OUTPUT
VOLTAGE RANGE |
± 70
V |
| OUTPUT
BANDWIDTH |
10
kHz (measured across 1K load R) |
| INPUT
RESISTANCE |
>
20 M Ohms |
| INPUT
VOLTAGE @ Imax |
± 10
volts |
| INPUT/OUTPUT
LINEARITY ERROR |
<
0.5% |
| Rise, Fall Time |
26 µs @ 10 K |
| POWER |
|
Model
A395D
|
17
alkaline 9 V batteries |
|
Model
A395R
|
17
rechargeable 9 V batteries |
| DIMENSIONS |
6.5 x
4 x 3.5 in. (16 x 10 x 9 cm) |
| SHIPPING
WEIGHT |
4 lb
(1.8 kg) |
|
|
 |
| Manual  |
 |
 |
 |
 |
 |
 |
 |
|
 |