| The term
stimulation refers to the delivery of energy of some kind to
a biological tissue in order to elicit an observable
response. Although the energy used in stimulation may be
chemical, thermal, mechanical or electrical, this discussion
will focus on electrical stimulation. Electrical stimulation
of biological tissues involves the delivery of current and
voltage to the stimulation site. The two quantities are
related by Ohm's law:
V=IR
where V is the applied voltage, I is the current and R is
the electrical resistance of the tissue and or the
stimulating electrodes. This simple equation shows that if
voltage is constant, current flow will diminish if the
tissue/electrode resistance goes up and will increase if the
resistance decreases. More commonly, the resistance of
tissue differs from sample to sample, and the resistance of
the electrodes changes with applied current over time in a
process called polarization.
Types of Stimulus Devices
In the delivery of electrical energy to biological
tissues, stimulus devices can hold either current or voltage
constant during the stimulating process. Devices that hold
voltage constant at a value set by the user and allow
current to be determined by Ohm's law are known as constant
voltage stimulators. Devices that hold current constant at a
value set by the user during the stimulation process and
allow voltage to be determined by Ohm's law are called
constant current stimulators. Constant current stimulators
are preferred for two reasons:
- First, current is the quantity that stimulates most
excitable tissues.
- Second, stimulating electrodes tend to increase
their resistance as stimulation progresses, as do some
tissues. A constant current stimulator will “sense”
resistance change and provide whatever voltage is needed
to maintain delivery of current at the set rate.
There is obviously a limit to how much voltage a constant
current stimulator can provide. If the resistance of the
preparation were to become infinite, as might happen if one
of the stimulating electrodes was removed from the tissue,
the stimulator could not mount an infinite voltage to
compensate. The maximum amount of voltage that a constant
current stimulator can provide is called the compliance
voltage. Once this compliance voltage has been reached,
further increases in tissue resistance will cause a drop in
delivered current. WPI isolators in the A360 and DLS series
offer a compliance voltage of 100V with very low noise.
Stimulus isolators, as the name implies, also isolate a
given stimulus from ground. In an instrument design context
most people think of isolation from ground as it relates to
electrical safety. From a biological recording standpoint
there are other issues. Consider the circuit in Figure 1.

Fig.1
The stimulator in this case is a battery with a switch;
current leaves the positive terminal of the battery, travels
down the stimulating lead, passes through the tissue and
returns to the negative terminal of the battery. 100% of
the current delivered returns to the negative terminal of
the battery. The figure also shows a voltmeter in the
preparation. Voltages generated by the tissue as a
consequence of the stimulus are recorded with respect to the
voltmeter’s ground electrode. In Figure 2 the
battery is replaced with a line powered stimulator. Even
though the stimulus source and the voltmeter have separate
ground electrodes, they represent the same electrical point.
For this reason a significant portion of the stimulus
current returns to ground by way of the voltmeter’s ground
lead. If the currents are significant, or the voltages that
you are trying to measure are very small, the I x R drop
across the resistance of the voltmeter’s ground electrode
will add to the recorded voltage from the tissue and will be
seen as a DC artefact.

Fig.2 Capacitative coupling between the voltmeter
circuit and the isolated circuit can induce current to flow
in the voltmeter ground. The induced current will
transiently flow across the resistance of the voltmeter
ground and its I x R drop will be seen as a transient spike
in front of and behind any pulse of current delivered by the
current source. This is termed an AC artefact. The primary
reason why researchers use an isolated current source is to
minimize artefact. But what if you are not recording? There
is no voltmeter ground to produce an observable artefact.
If you stop and consider the examples above, the artefact
was minimized or eliminated by controlling the path of the
stimulus current. Knowing the current path can be critical
in physiological stimulation.
Consider the current path in Figure 3. The animal
represented by the badly drawn cat is secured in a
stereotaxic frame. The frame is grounded. The animal
contacts the frame at multiple points. Figure 3 shows the
battery and switch model of a stimulator. Here as before all
of the current that leaves the positive side of the battery
must return to the negative side. 100% of the stimulus
current must pass between the stimulus electrodes. In this
case the path as well as the exact amount of current
delivered is known.

Fig. 3 Figure 4 shows the same experimental setup
except the battery has been replaced by a line powered
voltage source. Stimulus current now returns to ground via
many routes. The amount of current that flows to ground is
determined by the resistance between the source and each of
the ground points and is calculated as a resistive network
using Kirtchoff's laws. Many unintended areas of the animal
may be stimulated.

Fig. 4 In the real world we cannot use
a battery and a switch, particularly if the current
durations are on the order of milliseconds. Electronic
devices such as pulse generators and computers are used to
generate timing, and Isolators driven by these devices are
used to deliver the stimulus. By connecting an isolator
(even a battery powered one) to a mains powered pulse
generator, you connect the isolated ground of the isolator
to the mains ground of the pulse generator. Unless the
electrical connection between the two devices is
accomplished without using a mechanical connection between
the two devices you break the isolation. This is what makes
an isolator an isolator. The non-mechanical contact between
machines that constitutes the isolation barrier can be
accomplished in one of two ways. Originally,
isolators were transformer isolated: pulse waves were
applied to the primary winding of a transformer, while the
actual stimulus was derived from the secondary winding. The
transference was accomplished by induction. This approach
suffered from two shortcomings. The devices could not pass
DC so no constant voltage isolated stimulations could be
made. The transformer approach also has intrinsically higher
capacitance. This means that while the resistance between
the primary and secondary coils is very high, the high
capacitance will produce an AC artefact that is unacceptably
large compared to other isolation techniques.
Optical isolation is the second popular scheme. The field
has all but standardized on this. In simple terms, the
input pulse wave powers a light that shines across the
barrier onto a photocell that produces the stimulus wave.
There have been countless variations on this theme and it
is used for isolating recording amplifiers as well as
stimulators. All of WPI’s isolators are optical. |