Vibroslice
uses a vibrating blade to slice tissues without the trauma
produced by other methods. Live brain or other tissues can
be cut into slices 50- to 700-microns thick. Fixed tissues
can be cut down to 20-micron slices (these need not be
embedded or frozen). Particularly useful for improving the
access for certain histological reagents (e.g., during
processing for horseradish peroxidase).
The
blade has a lateral displacement of about 1 mm, and
its oscillating frequency may be varied between 60 and
4500 rpm. This allows you to achieve clean cuts in tissues
of different mechanical consistencies.
Model
NVSL offers a manual advance for positioning the specimen
holder and bath chamber. Sample positioning on Model VSLM1
is motorized. Other features include independent,
removable specimen holder and bath chamber, variable
advance speed (motorized 0.07 - 1.6 mm/s version only) and
hands-free operation via a footswitch.
|
SPECIFICATIONS
|
| SECTION
THICKNESS |
|
MINIMUM
|
20
µm (fixed tissue) 50 µm (fresh tissue) |
|
MAXIMUM
|
700
µm |
|
STEP
SIZE
|
5
µm |
| CHAMBER |
|
DIMENSIONS
|
21/8
x 31/4 x 11/4 in. (55 x 82 x 31 mm) |
|
VOLUME
|
8.5
cu. in. (140 ml) |
| BATH |
|
ADVANCE
SPEED
|
0.07
to 1.6 mm/s (VSLM1) (continuously variable) |
|
REVERSE
SPEED
|
4
mm/s (VSLM1) |
| VIBRATION |
|
SPEED
|
60
to 4500 rpm |
|
AMPLITUDE
|
1
mm (approx.) |
| BLADE
ANGLE |
Variable
through 90° from horizontal to vertical |
| POWER |
95-135
V, 50/60 Hz, or 220-240 V, 50/60 Hz |
| SHIPPING
WEIGHT |
19
lb (8.6 kg) |
IntegraSlice
The
unprecedented quality of slice of IntegraSlice stems from
its technical design features including cutting head
mounted upon a parallel leaf-spring assuring blade travel
along a line and arcuate path of the blade avoiding static
contact of the blade with the advancing tissue.
Sophisticated control allows the advance speed to be
varied from minus 1.00mm/sec and up to 3.00mm/sec in steps
of 0.01mm. During blade return the machine automatically
lowers the tissue away from the blade and then readjusts
to the original height once blade retraction has been
completed. In all operations, even at the slowest rates,
blade advance and blade speed is accurately maintained
exactly as the user has specified due to constant feedback
control. Slice thickness can be incremented in 1 micron
steps and controlled to a maximum speed of 1 micron per
second.
The
unprecedented quality of slice of IntegraSlice stems from
its technical design features including cutting head
mounted upon a parallel leaf-spring assuring blade travel
along a line and arcuate path of the blade avoiding static
contact of the blade with the advancing tissue.
Sophisticated control allows the advance speed to be
varied from minus 1.00mm/sec and up to 3.00mm/sec in steps
of 0.01mm. During blade return the machine automatically
lowers the tissue away from the blade and then readjusts
to the original height once blade retraction has been
completed. In all operations, even at the slowest rates,
blade advance and blade speed is accurately maintained
exactly as the user has specified due to constant feedback
control. Slice thickness can be incremented in 1 micron
steps and controlled to a maximum speed of 1 micron per
second.
Two Models Available:
IntegraSlice-P—
Fully programmable cutting profiles of speed,
distance, section thickness
-
Dynamically Adjustable Speed and Amplitude (DASA)
- both the speed and the amplitude of the cut can be
dynamically adjusted, whilst the blade is in motion
-
Instrument “learns” manually controlled cycle,
which can be repeated automatically
-
Cutting profiles can be set and stored
IntegraSlice-M—
Motorized or Manual advance of the oscillating
blade
IntegraSlice
design combines accuracy, speed and the flexibility to
remove the tissue bath and mounting for transport of the
tissue, for cleaning and autoclaving. integraslice is
supplied with a standard tissue bath and holder, which can
be replaced by the autoclavable version or the temperature
controlled version.
IntegraSlice is controlled using sealed Liquid Crystal
Display (LCD) with associated membrane keypad to set the
cutting parameters of blade speed, advance speed and
section thickness. There is a rotary brightness control
for the cold light source. The IntegraSlice operation is
commenced by actuating the two-stage footswitch. The first
stage starts the oscillating blade and the second stage
starts the advance of the blade.
|
SPECIFICATIONS
|
|
Blade
Motion
|
Arcuate
with minimal vertical movement
|
|
Oscillation
Speed
|
Linear
from 20-120 Hz
|
|
Amplitude
of Cut
|
Nominally
0.5 to 1.5 mm
|
|
Blade
Advance Speed
|
Between
minus 1.00 mm to plus 3.00 mm/sec in 0.01 mm steps
|
|
Blade
Return Speed
|
4.0
mm/sec whilst tissue is automatically lowered from blade
|
|
Vertical
Range
|
32
mm total, in 1 µm steps or maximum speed of 1 mm/sec
|
|
Sectioning
Range
|
1-40
mm
|
|
Maximum
Size of Specimen
|
30
x 40 mm
|
|
Tissue
Bath options
|
Standard,
Autoclavable, or Temperature Controlled
|
|
Cutting
Head Options
|
Standard
or autoclavable
|
|
Blades
|
Stainless
steel or ceramic single bevel
|
|
Observation
|
Inspection
Microscope 10x-40x with zoom and Clearview window
|
|
Voltage
Input
|
115/230
Volts
|
|
Rated
Frequency
|
50-60
Hz
|
|