There are 9 images available to view - click to enlarge and scroll through the product gallery.
BEV-1000 Instruction Manual
/ Download as PDF
The BEV-1000 combines high-magnification imaging and precise mechanical control to support accurate beveling across a wide range of glass capillary sizes, as outlined
below:
- Total magnification (optical + digital): 230×
- 9” TFT panel display resolution: 1024 × 768
- Beveling operation speed: 0 - 3,000 RPM
- Flat diamond wheel with a surface variance of <2 μm/cm ensure you get a clean bevel with less tip breakage
- Compatible with range of glass capillary sizes, up to 3 mm outer diameter
- Excellent flexibility and control over the finished beveled angle
Benefits
- Large LCD screen for enhanced visibility, and reduced eye strain vs relying on microscope eyepieces for visualization
- Adjustable speed and brightness all on one plane for easy adjustment during use
- Protective tinted shield both defends against residual glass debris during beveling and brightness from light source
- All-in-one system, all required components are included, no hidden costs or add-ons needed
- Straightforward axis manipulation for pipette vs camera adjustment
Specifications
| Type | Description |
| Camera Magnification | 230× (total optical and digital) |
| Camera Working Distance (or focal length | > 12 mm |
| Camera Field of View | > 0.8 mm |
| Manipulation of Camera | |
|
7 mm |
|
8 mm |
| Speed of Abrasive Diamond Wheel | 0 – 3000 RPM |
| Abrasive Diamond Wheel Surface Flatness | < 2 µm/cm |
| Adaptable Glass Tube Range | Ø1 – 3 mm OD |
| Pre-Pulled Glass Capillary Diameter Range | 0.5 to 100 µm* |
| Reticle Scale Resolution | 8 µm/division |
| Minimum Measurable Tip Size | 8 µm** |
| Minimum/Maximum Tip Beveling Range | 0.1 – 100 µm |
| Beveled Pre-Pulled Glass Capillary Diameter | < 200 µm |
| Manipulation of Glass Tube | |
|
10 mm |
|
10 mm |
|
0 – 180° |
|
0 – 15° |
| Sliding Track and Working Platform Inclination Angle | 30° |
| Display Size/Resolution | 9" TFT panel / 1024 × 768 |
| Display Screen Supply Voltage | 12 VDC |
| Backlight Voltage, Power | 12 VDC, 2 W |
| Mains Voltage, Frequency, Power | 110–240 VAC, 50/60 Hz, 62 W |
| Dimensions | 30 × 20 × 46 cm (L × W × H) |
| Net Weight | 10 kg (22 lbs.) |