How to Take Voltage Measurements with the EVOM™ Manual

How to Take Voltage Measurements with the EVOM™ Manual


EVOM™ Manual is WPI’s newest instrument to measure Trans Epithelial Electrical Resistance (TEER). Let’s look at how to set it up and take voltage measurements.

  1. Power on the EVOM™ Manual using the power switch on the rear panel.
  2. Insert the USB flash drive provided with the system into the USB port on the side of the meter.
  3. By default, the filename is plate1. You may choose a prefix that auto-indexes or a unique file name:
    • Press and hold the Store button on the main screen for 2 seconds or press Store Screen on the Setup menu to configure your stored data files.
    • To use a prefix and auto-indexing, activate the Auto Index radio button. Auto indexing sets the name to a numeric sequence with the name, a prefix and a sequential number (name_prefixN). Then, press Store Prefix to enter a file name. Or, to use a unique file name, deactivate the Auto Index radio button. Then, select File Name to enter a new name.
    • Press Clear.
    • Use the keypad to enter a new name.
    • Press Enter.
  4. Press Setup to access the Setup menu.
    • Select Plate–Choose you plate size.
    • Select Mode Units – Choose volts or millivolts (the normal mode is millivolts).
    • Electrode Equilibration – Zero the electrodes' galvanic charge in saline or media. The electrode needs to be equilibrated in the liquid, such as media, for a few hours (>3 Hours) to achieve a stable voltage.
    • Next, press the Probe Null button. Push the NULL button. Wait for the second message stating “Solution value 0.0” before returning. This action subtracts the offset voltage of the electrodes. Then activate the Enable/Disable radio button to offset the electrode potential.
    • Press Return to navigate back to the main menu.
  5. Begin measuring by pressing the foot switch. Or, touch the Store button on the screen and tap the Next Well area on the right side of the screen to advance to the next well. Repeat the sampling process.
  6. Once the last well has been recorded, the file storing selection notification appears.
  7. Press Store. The EVOM™ Manual should respond with “File saved” within 15 seconds. If “Saving file...” is still seen after 2 minutes, press the Cancel button and try a different USB drive. Once a file has been saved, open the file on a computer to verify the contents.

Once you get the system setup, taking measurements is easy.


EVM-MT-03-01

EVM-MT-03-01

EVOM™ Manual Epithelial Volt Ohm Meter for TEER Measurement

View details...

Our Clients Include:

GlaxoSmithKline
University College London
Novartis
Imperial College
University of Cambridge
University of Oxford

Keep in Touch

We promise NEVER to share your details with anyone. You can opt out at any time.